Anal Acoustic Reflectometry: A New Reproducible Technique Providing Physiological Assessment of Anal Sphincter Function

作者:Mitchell Peter J*; Klarskov Niels; Telford Karen J; Hosker Gordon L; Lose Gunnar; Kiff Edward S
来源:Diseases of the Colon & Rectum, 2011, 54(9): 1122-1128.
DOI:10.1097/DCR.0b013e318223fbcb

摘要

BACKGROUND: Anal acoustic reflectometry is a new technique of assessing anal sphincter function. Five new variables reflecting anal canal function are measured: the opening and closing pressure, the opening and closing elastance, and hysteresis.
OBJECTIVE: This study aimed to compare the reproducibility of this new technique, in terms of test-retest and interrater reliability, with manometry, the current standard test of sphincter function.
DESIGN: This is a comparative study of reproducibility between anal acoustic reflectometry and manometry.
SETTINGS: This study was conducted at a university hospital (outpatient clinic and endoscopy unit).
PATIENTS: Twenty-six (21 female) subjects were assessed with both anal acoustic reflectometry and manometry on 2 separate occasions (test-retest reliability) and 22 (16 female) subjects were assessed with both methods by 2 separate investigators (interrater reliability).
MAIN OUTCOME MEASURES: Reproducibility was assessed according to the Bland-Altman method.
RESULTS: All of the measured novel anal acoustic reflectometry parameters had acceptable mean differences and repeatability coefficients. Comparison of the 2 methods of sphincter assessment (anal acoustic reflectometry vs manometry) was made for measurements taken at rest and during voluntary contraction. There was no significant difference in terms of test-retest reliability between the manometry maximum resting pressure vs the reflectometry opening pressure (P = .57) or manometry maximum squeeze pressure vs the reflectometry squeeze opening pressure (P = .68). No significant difference between methods was found in interrater reliability during assessments at rest (P = .62) and voluntary contraction (P = .96).
LIMITATIONS: Anal acoustic reflectometry is limited, as with all tests of anorectal function, in that the device is placed within the anal canal, causing stimulation of sensory and stretch receptors.
CONCLUSIONS: Anal acoustic reflectometry has a reproducibility comparable to manometry in terms of both test-retest and interrater reliability. Anal acoustic reflectometry is a promising technique that allows an assessment of anal canal physiology that is not available with manometry.

  • 出版日期2011-9