摘要

A procedure employing inductively coupled plasma atomic emission spectroscopy has been developed for the precise determination of the composition of Zn1+xGe1+xP2+x alloys after their autoclave digestion. An optimal composition of the acid mixture has been chosen and the conditions preventing element losses in the digestion process have been found. To minimize the error of the developed procedure, the optimal concentration of Zn, Ge, and P in the analytes was found by calculating the results of analysis as atomic ratios Zn : Ge and P : Ge. The accuracy of the developed procedure was confirmed by the standard addition method and the comparison of the results with the results of an independent method. The relative standard deviation of the determination of atomic ratios was 0. 7-2.0%.

  • 出版日期2011-6