Background. Standard non-invasive arterial pressure (AP) measurements are discontinuous. By providing non-invasive beat-to-beat AP measurements, Nexfin (TM) might limit duration of intraoperative hypotension and hypertension. We assessed the ability of Nexfin (TM) to detect AP variations by comparing its trending ability with invasive AP monitoring. %26lt;br%26gt;Methods. Thirty-one subjects undergoing elective surgery under general anaesthesia were included. During induction, simultaneous pairs of AP measurements were collected every 5 s from the Nexfin (TM) finger sensor and a homolateral radial artery catheter. Magnitude and time lags of AP variations from baseline to nadir and peak were calculated for both methods. Concordance analysis was performed by the Bland-Altman method (for comparison of repeated measures when appropriate). %26lt;br%26gt;Results. Nexfin (TM) detected 100% of AP changes with the median delays of Os (-13 to 7) and Os (-5 to 12) for nadir and peak, respectively. Bias [limits of agreement (LOA)] of systolic AP (SAP) variations was -0.5 mm Hg (-31.2 to 30.2) and -9.4 mm Hg (-31.3 to 12.6) from baseline to nadir and from baseline to peak, respectively. For 3479 analysed paired measurements, bias was -3.8 and -8.8 mm Hg for SAP and diastolic AP, with LOA of (-36.0 to 28.5) and (-29.8 to 12.3), respectively. %26lt;br%26gt;Conclusions. Nexfin (TM) detects AP variations accurately and can be a useful warning device during anaesthesia. However, it is not interchangeable with invasive monitoring, given the large LOA between the two measurements.

  • 出版日期2014-7