Accuracy of image-free navigation for severely deformed knees

作者:Takasaki Minoru; Matsuda Shuichi*; Fukagawa Shingo; Mitsuyasu Hiroaki; Miura Hiromasa; Iwamoto Yukihide
来源:Knee Surgery, Sports Traumatology, Arthroscopy, 2010, 18(6): 763-768.
DOI:10.1007/s00167-009-0944-4

摘要

This study evaluated the accuracy of an image-free navigated total knee arthroplasty (TKA) system when used to align deformed tibia bone models. The accuracy was assessed in normal, 10A degrees varus, 20A degrees varus, 10A degrees valgus, and 20A degrees valgus tibia bone models (a total of five tibial models) by direct measurement of the navigated cutting guide. The mean angular errors in the tibial mechanical axes of the normal, 10A degrees, and 20A degrees varus models, respectively, were 0.0A degrees, 0.7A degrees varus, and 2.4A degrees varus. Thus, the errors seen with the two varus models were significantly larger than that associated with the normal model. The mean angular errors were 0.1A degrees varus and 0.4A degrees valgus in the 10A degrees and 20A degrees valgus models, respectively. These errors were not significantly different from those obtained with the normal model. These results suggest that in varus-deformed knees, image-free navigation has a tendency to cut the tibia in varus. This fact is considered to be one of the reasons for the lack of superiority of TKA alignment in severely deformed knees when using image-free navigation. Therefore, special attention must be paid when using image-free navigation TKA in such cases.

  • 出版日期2010-6