The electrooxidative behavior and determination of zolmitriptan at a boron-doped diamond electrode were investigated using cyclic, linear sweep, differential pulse and square wave voltammetric techniques. Zolmitriptan undergoes irreversible oxidation at a peak potential of about +0.9 V (vs Ag/AgCl/3 M KCl). DPV and SWV techniques are proposed for the determination of zolmitriptan in phosphate buffer at pH 3.03, which allows quantitation over the two different ranges (8 x 10(-7)-8 x 10(-6) M and 1 x 10(-5)-1 x 10(-4) M) in supporting electrolyte for both methods. A linear response was obtained in phosphate buffer over two different ranges (6 x 10(-7)-8 x 10(-6) M and 1 x 10(-5)-1 x 10(-4) M) for spiked serum samples at pH 3.03 for both techniques. The repeatability and reproducibility of the methods for all media were determined. The standard addition method was used in serum. Precision and accuracy were also checked in all media. No electroactive interferences from the excipients and endegenous substances were found in the pharmaceutical dosage form and the biological sample, respectively.