Purpose: To investigate the outcomes of patients who were treated for thoracoabdominal aortic aneurysms (TAAAs) using custom-made fenestrated-branched stent-grafts. Methods: A consecutive series of 108 patients (mean age 73.5 years; 73 men) with TAAA were treated with E-xtra Design Engineering customized fenestrated-branched stent-grafts between November 2011 and January 2017. Data on baseline characteristics, procedures, and clinical follow-up were collected from 6 regional European surgical centers for retrospective analysis of endoleaks, reinterventions, and target vessel patency. The median aneurysm diameter was 6.75 cm (range 5.5-13). The distribution of the TAAA according to the modified Crawford classification of extent was 25 (24%) type I, 19 (17%) type II, 20 (18%) type III, 29 (27%) type IV, and 15 (14%) type V. Results: Technical success was achieved in 95% (103/108) of cases. Major early perioperative complications occurred in 40 (37%) patients. The 30-day mortality was 9.2% (10/108), and perioperative spinal cord ischemia was observed in 6 (5.5%) patients [2 (1.8%) permanent]. During the mean follow-up of 17.6 months (range 3-52), 28 (26%) patients required late reintervention. Two patients died due to aneurysm- or procedure-related causes. The estimated survival rates at 1, 2, and 4 years were 87%, 84%, and 51%, respectively. The estimated target vessel patency rates at the same time points were 95%, 91%, and 90%, respectively. The freedom from reintervention estimates were 84% and 73% at 1 and 4 years, respectively. Conclusion: Endovascular repair of TAAA using Jotec customized fenestrated-branched stent-grafts appears to be safe and effective in the early to midterm. The considerable rate of secondary interventions indicates that further improvements, graft surveillance, and follow-up are required.

  • 出版日期2018-4