A fully implantable telemetry system for the long-term measurement of habitual bone strain

作者:de Jong W C*; Koolstra J H; van Ruijven L J; Korfage J A M; Langenbach G E J
来源:Journal of Biomechanics, 2010, 43(3): 587-591.
DOI:10.1016/j.jbiomech.2009.09.036

摘要

Long-term in-vivo recordings of habitual bone strain in freely moving animals are needed to better understand the everyday mechanical loading environment responsible for bone-tissue maintenance. However, wireless methods to make such recordings are scarce. We report on the successful customisation of a commercially available voltage transmitter hooked-LIP to a strain-gauge rosette, its subcutaneous implantation in rabbits, and the quality of the implant's strain-gauge recordings. Continuous wireless recordings of a completely operational strain-gauge rosette glued to the mandibular surface of a freely moving rabbit could be made up to 33 h. The resolution of the system was 1.5 microstrains/bit. The noise in the signal was 4.5 microstrains. To facilitate the automatic Counting of bone-strain events in the retrieved data, and to calculate their peak amplitude, a novel approach is presented. The described technique enables the quantification of the daily bone-strain history defining the architecture and composition of bone tissue, and can help to further elucidate the strain parameters which influence bone tissue.