摘要

Contrast-enhanced C-arm CT is routinely used for intra-operative guidance during the trans-catheter aortic valve implantation (TAVI); however, the requirement for contrast agent injection is not preferable, especially for patients with renal insufficiencies. To address this problem, we present a novel framework for fully automatic registration of pre-operative CT and non-contrast-enhanced C-arm CT. The proposed framework provides an improved workflow and minimizes the usage of contrast agent in the TAVI procedure. Our framework consists of three steps: coarse rigid-body alignment, anatomical knowledge-based prior deformation field generation, and fine deformable registration. We validated the proposed framework on 20 real patient data sets. Based on the 20 data sets, the mesh-to-mesh errors at the aortic root from different methods are measured. Our proposed method significantly outperforms the other state-of-the-art methods. Specifically, we achieve the registration accuracy at 1.76 +/- 0.43 mm which is clinically plausible. Quantitative evaluation on real non-contrast enhanced C-arm CT data sets confirms the applicability in the clinical usage. The proposed heart registration method is generic and hence can be easily applied to other cardiac applications.

  • 出版日期2014-12