Dual-band Fourier domain optical coherence tomography with depth-related compensations

作者:Zhang Miao*; Ma Lixin; Yu Ping
来源:Biomedical Optics Express, 2014, 5(1): 167-182.
DOI:10.1364/BOE.5.5.000167

摘要

Dual-band Fourier domain optical coherence tomography (FDOCT) provides depth-resolved spectroscopic imaging that enhances tissue contrast and reduces image speckle. However, previous dual-band FD-OCT systems could not correctly give the tissue spectroscopic contrast due to depth-related discrepancy in the imaging method and attenuation in biological tissue samples. We designed a new dual-band full-range FDOCT imaging system and developed an algorithm to compensate depth-related fall-off and light attenuation. In our imaging system, the images from two wavelength bands were intrinsically overlapped and their intensities were balanced. The processing time of dual-band OCT image reconstruction and depth-related compensations were minimized by using multiple threads that execute in parallel. Using the newly developed system, we studied tissue phantoms and human cancer xenografts and muscle tissues dissected from severely compromised immune deficient mice. Improved spectroscopic contrast and sensitivity were achieved, benefiting from the depth-related compensations.

  • 出版日期2014-1-1

全文