Absolute calibration for Fizeau interferometer with the global optimized shift-rotation method

作者:Song Weihong*; Li Shifang; Hou Xi; Wu Fan; Zhao Wenchuan; Wan Yongjian
来源:Optics and Lasers in Engineering, 2014, 54: 49-54.
DOI:10.1016/j.optlaseng.2013.10.005

摘要

We present an absolute calibration method for Fizeau interferometer with global optimized shift-rotation, which requires rotational and translational measurements. With a data matching method to decrease the alignment error of rotations and translations, the accuracy can be guaranteed. The global optimized shift-rotation method can obtain the surface deviation accurately of both the reference and test surfaces in terms of Zernike polynomials. Experiments have been implemented to verify the validity of the absolute calibration, and the results show that the reference surface deviation agrees well with the calibration result of conventional two-sphere method. Simulations were performed to evaluate the errors introduced by rotations and translations.