This in vitro investigation was conducted to study the relationship between resonance frequency analysis (RFA) and lateral displacement measurements of dental implants. A total of 30 implant sites were prepared in nine fresh bovine bone specimens. The bone density around each preparation was determined by using cone beam computerized tomography (CBCT) and imaging software. Dental implants were then inserted during continuous registration of insertion torque. RFA measurements were performed in perpendicular and parallel to the long axis of the specimens. The bone blocks were embedded in plaster and fixated in a specially designed rig for displacement measurements. A lateral force of 25N was applied via an abutment perpendicular and parallel to each implant and the displacement measured in m. In addition, a flex constant (m N1) was calculated for each measurement. There was a significant inverse correlation between RFA and lateral implant displacement (m) measurements and between RFA measurements and the flex constant in both perpendicular and parallel directions in bone (P0 center dot 001). Moreover, both RFA and displacement measurements correlated with bone density (P0 center dot 001). It is concluded that RFA measurements reflect the micromobility of dental implants, which in turn is determined by the bone density at the implant site.

  • 出版日期2013-3