摘要

A carbon screen-printed electrode modified insitu with lead film (PbF-SPCE) was applied for the adsorptive stripping voltammetric determination of Co(II) in the form of a complex with 1,2-cyclohexanedione dioxime. Lead film was electrochemically deposited in situ on SPCE from a 0.2M ammonia buffer solution (pH 8.7) containing 5.10(-5) M Pb(NO3)(2) and 5.10(-5) M nioxime. Due to the very low LOD (0.003 mu g L-1, i.e., 0.05 nmol L-1 Co(II); t(acc)=120s), the developed procedure could be rated among the most sensitive methods employing SPEs. The Ni(II) signal was significantly lower than the Co(II) one and the separation of Ni(II) and Co(II) peaks was even better at the PbF-SPCE than at the hanging mercury drop electrode.

  • 出版日期2016-2