Battery operated preconcentration-assisted lateral flow assay

作者:Kim Cheonjung; Yoo Yong Kyoung; Il Han Sung; Lee Junwoo; Lee Dohwan; Lee Kyungjae; Hwang Kyo Seon; Lee Kyu Hyoung; Chung Seok*; Lee Jeong Hoon*
来源:Lab on a Chip, 2017, 17(14): 2451-2458.
DOI:10.1039/c7lc00036g

摘要

Paper-based analytical devices (e.g. lateral flow assays) are highly advantageous as portable diagnostic systems owing to their low costs and ease of use. Because of their low sensitivity and detection limits for biomolecules, these devices have several limitations in applications for real-field diagnosis. Here, we demonstrate a paper-based preconcentration enhanced lateral flow assay using a commercial beta-hCG-based test. Utilizing a simple 9 V battery operation with a low power consumption of approximately 81 mu W, we acquire a 25-fold preconcentration factor, demonstrating a clear sensitivity enhancement in the colorimetric lateral flow assay; consequently, clear colors are observed in a rapid kit test line, which cannot be monitored without preconcentration. This device can also facilitate a semi-quantitative platform using the saturation value and/or color intensity in both paper-based colorimetric assays and smartphone-based diagnostics.

  • 出版日期2017-7-21