MRI Contrast From Relaxation Along a Fictitious Field (RAFF)

作者:Liimatainen Timo; Sorce Dennis J; O'Connell Robert; Garwood Michael; Michaeli Shalom*
来源:Magnetic Resonance in Medicine, 2010, 64(4): 983-994.
DOI:10.1002/mrm.22372

摘要

A new method to measure rotating frame relaxation and to create contrast for MRI is introduced. The technique exploits relaxation along a fictitious field (RAFF) generated by amplitude- and frequency-modulated irradiation in a subadiabatic condition. Here, RAFF is demonstrated using a radiofrequency pulse based on sine and cosine amplitude and frequency modulations of equal amplitudes, which gives rise to a stationary fictitious magnetic field in a doubly rotating frame. According to dipolar relaxation theory, the RAFF relaxation time constant (T(RAFF)) was found to differ from laboratory frame relaxation times (T(1) and T(2)) and rotating frame relaxation times (T(1p) and T(2p)). This prediction was supported by experimental results obtained from human brain in vivo and three different solutions. Results from relaxation mapping in human brain demonstrated the ability to create MRI contrast based on RAFF. The value of T(RAFF) was found to be insensitive to the initial orientation of the magnetization vector. In the RAFF method, the useful bandwidth did not decrease as the train length increased. Finally, as compared with an adiabatic pulse train of equal duration, RAFF required less radiofrequency power and therefore can be more readily used for rotating frame relaxation studies in humans. Magn Reson Med 64:983-994, 2010.

  • 出版日期2010-10