The presence of endocardial leads may limit applicability of coronary CT angiography

作者:Sosnowski Maciej; Mlynarski Rafal*; Wlodyka Agnieszka; Brzoska Jan; Kargul Wlodzimierz; Tendera Michal
来源:Scandinavian Cardiovascular Journal, 2010, 44(1): 31-36.
DOI:10.3109/14017430903114453

摘要

Objectives. Some indications that MSCT is necessary may be visible in patients with implanted pacemakers, CRTs and ICDs devices. Purpose. To evaluate whether the presence of endocardial leads influences the quality of visualization of the coronary arteries in MSCT. Design. In 37 patients a 64-slice MSCT (Aquilion 64, Toshiba) was performed due to a suspicion of ischemic heart disease (33 patients) or suspicion of lead perforation (four patients). The control group consists of 35 patients. Influence of the leads was evaluated independently by two cardiologists. Results. The total number of leads that were evaluated was 63. In 26 cases (41%) significant problems with the evaluation of coronary arteries occurred: LM-1 (2%) case, LAD-3 (5%) cases, LCx-4 (6%) cases, RCA-18 (28%) cases. The presence of a right ventricular lead did not influence the quality of the imaging of the left main coronary artery (LM) and its main branches (LAD and LCx). The main concerns were related to the RCA visualization-interpretable images in 18/63 (29%) of the analyzed leads. Conclusions. The presence of implanted endocardial leads may limit the applicability of coronary CT angiography in patients with implanted anti-arrhythmic devices.

  • 出版日期2010-2