Mid term results of total hip arthroplasty using polyethylene-ceramic composite (Sandwich) liner

作者:Wang, Tao; Sun, Jun-Ying*; Zha, Guo-Chun; Dong, Sheng-Jie; Zhao, Xi-Jiang
来源:Indian Journal of Orthopaedics, 2016, 50(1): 10-15.
DOI:10.4103/0019-5413.173512

摘要

Background: Ceramic-on-ceramic (COC) couplings are an attractive alternative bearing surfaces that have been reported to eliminate or reduce problems related to polyethylene wear debris. However, the material in total hip arthroplasty (THA) remains one of the major concern regarding the risk of fracture. The present study aims at reporting the fracture rate of bearings in a series of COC THAs with the use of a sandwich liner and attempt to detect the relative risk factors, the possible cause and assess the clinical results. Materials and Methods: We retrospectively evaluated 153 patients (163 hips) using the sandwich liner COC THA between 2001 and 2009. Patient assessment was based on demographic factors, including age, weight, gender and body-mass index (BMI). All patients were evaluated clinically and radiographically or using computed tomography viz-a-viz dislocation, osteolysis, periprosthetic fracture, infection, loosening and implant fracture. Results: Three ceramic sandwich liners fracture (1.84%) were observed at an average of 7.3 years' followup. The factors which were found to be non-significant to the ceramic liner fracture, included age (P = 0.205), weight (P = 0.241), gender (P = 0.553), BMI (P = 0.736), inclination (P = 0.199) and anteversion (P = 0.223). The overall survival was 91.4% at 12-year with revision as the endpoint. Other complications included osteolysis in 4 (2.45%), dislocation in one and periprosthetic fracture in one. In no hip aseptic loosening of the implants was seen. Conclusions: Our experience with the ceramic-polyethylene sandwich liner acetabular component has been disappointing because of the high rate of fracture and osteolyis. We have discontinued the use of this device and recommend the same.