In this current study, the molten salt aerosol-laser-induced breakdown spectroscopy (LIBS) system was used to measure the uranium (U) content in a ternary UCl3-LiCl-KCl salt to investigate and assess a near real-time analytical approach for material safeguards and accountability. Experiments were conducted using five different U concentrations to determine the analytical figures of merit for the system with respect to U. In the analysis, three U lines were used to develop univariate calibration curves at the 367.01nm, 385.96nm, and 387.10nm lines. The 367.01nm line had the lowest limit of detection (LOD) of 0.065 wt% U. The 385.96nm line had the best root mean square error of cross-validation (RMSECV) of 0.20 wt% U. In addition to the univariate calibration approach, a multivariate partial least squares (PLS) model was developed to further analyze the data. Using partial least squares (PLS) modeling, an RMSECV of 0.085 wt% U was determined. The RMSECV from the multivariate approach was significantly better than the univariate case and the PLS model is recommended for future LIBS analysis. Overall, the aerosol-LIBS system performed well in monitoring the U concentration and it is expected that the system could be used to quantitatively determine the U compositions within the normal operational concentrations of U in pyroprocessing molten salts.