Gram-scale cryogenic calorimeters for rare-event searches

作者:Strauss R*; Rothe J; Angloher G; Bento A; Guetlein A; Hauff D; Kluck H; Mancuso M; Oberauer L; Petricca F; Proebst F; Schieck J; Schoenert S; Seidel W; Stodolsky L
来源:PHYSICAL REVIEW D, 2017, 96(2): 022009.
DOI:10.1103/PhysRevD.96.022009

摘要

The energy threshold of a cryogenic calorimeter can be lowered by reducing its size. This is of importance since the resulting increase in signal rate enables new approaches in rare-event searches, including the detection of MeV mass dark matter and coherent scattering of reactor or solar neutrinos. A scaling law for energy threshold vs detector size is given. We analyze the possibility of lowering the threshold of a gram-scale cryogenic calorimeter to the few eV regime. A prototype 0.5 g Al2O3 device achieved an energy threshold of E-th = (19.7 +/- 0.9) eV, the lowest value reported for a macroscopic calorimeter.

  • 出版日期2017-7-28