摘要

Interfacing experiments to a clinical magnetic resonance imaging scanner which require synchronization with an imaging pulse sequence can be challenging unless access to a trigger-out is available. We present a simple, alternative approach to synchronize an experiment to a scanner using an external trigger to drive the scanner through the peripheral optical pulse rate monitor and cardiac trigger available on most clinical imagers. A trigger circuit and pulse rate monitor interface are described for applying a stimulus in the form of a pulsed voltage to a sample at a specific time in a spin-echo, echo planar imaging sequence. The apparatus and approach could be used for many other types and numbers of experimental stimuli or events.

  • 出版日期2014-4