Digital Heart-Rate Variability Parameter Monitoring and Assessment ASIC

作者:Massagram Wansuree*; Hafner Noah; Chen Mingqi; Macchiarulo Luca; Lubecke Victor M; Boric Lubecke Olga
来源:IEEE Transactions on Biomedical Circuits and Systems, 2010, 4(1): 19-26.
DOI:10.1109/TBCAS.2009.2035555

摘要

This paper describes experimental results for an application-specific integrated circuit (ASIC), designed for digital heart rate variability (HRV) parameter monitoring and assessment. This ASIC chip measures beat-to-beat (RR) intervals and stores HRV parameters into its internal memory in real time. A wide range of short-term and long-term ECG signals obtained from Physionet was used for testing. The system detects R peaks with millisecond accuracy, and stores up to 2 min of continuous RR interval data and up to 4 min of RR interval histogram. The prototype chip was fabricated in a 0.5-mu m complementary metal-oxide semiconductor technology on a 3 x 3 mm(2) die area, with a measured dynamic power consumption of 10 mu W and measured leakage current of 2.62 nA. The HRV monitoring system including this HRV ASIC, an analog-to-digital converter, and a low complexity microcontroller was estimated to consume 32.5 mu V, which is seven times lower power than a stand-alone microcontroller performing the same functions. Compact size, low cost, and low power consumption make this chip suitable for a miniaturized portable HRV monitoring system.