摘要

Two novel cefquinome sulfate (CFQ)-selective electrodes were performed with dibutyl sebacate as a plasticizer using a polymeric matrix of polyvinyl chloride. Sensor 1 was prepared using sodium tetraphenylborate as a cation exchanger without incorporation of ionophore, whereas 2-hydroxy propyl beta-cyclodextrin was used as ionophore in sensor 2. A stable, reliable, and linear response was obtained in concentration ranges 3.2 x 10(-5) to 1 x10(-2) mol/L and 1 x 10(-5) to 1 x10(-2) mol/L for sensors 1 and 2, respectively. Both sensors could be sufficiently applied for quantitative determination of CFQ in the presence of degradation products either in bulk powder or in pharmaceutical formulations. Sensor 2 provided better selectivity and sensitivity, wider linearity range, and higher performance. Therefore it was used successfully for accurate determination of CFQ in biological fluids such as spiked plasma and milk samples. Furthermore, an online kinetic study was applied to the CFQ alkaline degradation process to estimate the reaction rate and half-life with feasible real-time monitoring. The developed sensors were found to be fast, accurate, sensitive, and precise compared with the manufacturer's reversed-phase chromatographic method.

  • 出版日期2016-2