Digital tomosynthesis (DTS) with a Circular X-ray tube: Its image reconstruction based on total-variation minimization and the image characteristics

作者:Park Y O*; Hong D K; Cho H S; Je U K; Oh J E; Lee M S; Kim H J; Lee S H; Jang W S; Cho H M; Choi S I; Koo Y S
来源:Journal of the Korean Physical Society, 2013, 63(5): 1060-1065.
DOI:10.3938/jkps.63.1060

摘要

In this paper, we introduce an effective imaging system for digital tomosynthesis (DTS) with a circular X-ray tube, the so-called circular-DTS (CDTS) system, and its image reconstruction algorithm based on the total-variation (TV) minimization method for low-dose, high-accuracy X-ray imaging. Here, the X-ray tube is equipped with a series of cathodes distributed around a rotating anode, and the detector remains stationary throughout the image acquisition. We considered a TV-based reconstruction algorithm that exploited the sparsity of the image with substantially high image accuracy. We implemented the algorithm for the CDTS geometry and successfully reconstructed images of high accuracy. The image characteristics were investigated quantitatively by using some figures of merit, including the universal-quality index (UQI) and the depth resolution. For selected tomographic angles of 20, 40, and 60A degrees, the corresponding UQI values in the tomographic view were estimated to be about 0.94, 0.97, and 0.98, and the depth resolutions were about 4.6, 3.1, and 1.2 voxels in full width at half maximum (FWHM), respectively. We expect the proposed method to be applicable to developing a next-generation dental or breast X-ray imaging system.

  • 出版日期2013-9

全文