摘要

Objective: A simple, reliable and specific method of HPLC which was performed on a reversed- phase Agilent SB column was developed for the determination of imatinib and CGP74588. @@@ Method: The mobile phase comprised water-acetonitrile-0.1% trifluoroacetic acid (59: 21: 20, v/v/v) flowed at the rate of 1 mL/min at 40 degrees C. The detection wavelength was 278nm. @@@ Results: Sensitivity has been illustrated in this method, and a good linear relationship was observed between 0.1-20.00 mu g/mL for imatinib with r(2) = 0.9994 and 0.01-2.00 mu g/mL for CGP74588 with r(2) = 0.9998 in rat plasma. The relative recoveries of imatinib and CGP74588 were all more than 93%, whilst the absolute recoveries of imatinib and CGP74588 were all more than 75%. Both intra-and inter-day precisions were all less than 4.5% for imatinib and 4.74% for GCP74588. The limits of quantification of these analytic methods were 0.1 mu g/mL for imatinib and 0.01 mu g/mL for CGP74588. @@@ Conclusion: The method was suitable for studying the pharmacokinetic parameters of imatinib and GCP74588 and routine analysis.