Assessment of MRI issues for the Argus II Retinal Prosthesis

作者:Weiland James D; Faraji Boozarjomehr; Greenberg Robert J; Humayun Mark S; Shellock Frank G*
来源:Magnetic Resonance Imaging, 2012, 20(3): 382-389.
DOI:10.1016/j.mri.2011.12.005

摘要

Objective: The objective was to evaluate magnetic resonance imaging (MRI) issues (magnetic field interactions, heating, artifacts and functional alterations) at 1.5 T and 3 T for the Argus II Retinal Prosthesis (Second Sight Medical Products, Sylmar, CA, USA).
Materials and Methods: Standardized protocols were used to assess magnetic field interactions (translational attraction and torque; 3 T, worst case), MRI-related heating (1.5 and 3 T), artifacts (3 T; worst case) and functional changes (1.5 and 3 T) associated with MRI.
Results: The magnetic field interactions were acceptable. MRI-related heating, which was studied at a relatively high, MR system-reported whole body averaged specific absorption rates, will not pose a hazard to the patient under the conditions used for testing. While artifacts were "moderate" in relation to the dimensions of the Argus II Retinal Prosthesis, optimization of MRI parameters can reduce the size of the artifacts. Exposures to MRI conditions at 1.5 and 3 T did not damage or alter the functional aspects of the Argus II Retinal Prosthesis.
Conclusions: In consideration of the test results, a patient with the Argus 11 Retinal Prosthesis may undergo MRI at 1.5 T or 3 T when specific guidelines and MRI conditions are followed, including those advised by the manufacturer.

  • 出版日期2012-4