ARDESIA Detection Module: A Four-Channel Array of SDDs for Mcps X-Ray Spectroscopy in Synchrotron Radiation Applications

作者:Bellotti G; Butt A D; Carminati M; Fiorini C*; Bombelli L; Borghi G; Piemonte C; Zorzi N; Balerna A
来源:IEEE Transactions on Nuclear Science, 2018, 65(7): 1355-1364.
DOI:10.1109/TNS.2018.2838673

摘要

The detection module of the ARDESIA spectrometer, an SDD-based, multichannel X-ray detector is presented. The detector design and experimental characterization are discussed. ARDESIA has been optimized for synchrotron radiation applications that require high-rate (>1-Mcps per channel) and high-resolution (better than 200 eV at 6 keV) X-ray detection. The main target applications considered in this development are X-ray fluorescence and X-ray absorption fine structure spectroscopy, in particular for the DAFNE-Light DXR1 beamline at LNF (Frascati, Italy). This detector is based on a compact unit (16 mm x 16 mm) that can he replicated and juxtaposed with several units, thus easily covering different detection configurations. The detector consists of a four-channel monolithic array of SDDs, produced by Fondazione Bruno Kessler (Trento, Italy) connected to a four-channel monolithic CUBE preamplifier mounted on the same carrier. The CUBE preamplifier, compatible with both analog and digital pulse processing, enables reaching high count rates (1-Mcps per channel), while keeping good spectroscopic performances. It is extremely low white series noise guarantees good energy resolution, even at short pulse processing times (below 200 ns).

  • 出版日期2018-7