A compact soft x-ray (0.1-1.2 keV) calibration bench for radiometric measurements using an original versatile Rowland circle grazing incidence monochromator

作者:Hubert S*
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2017, 855: 16-24.
DOI:10.1016/j.nima.2017.02.074

摘要

This paper describes an original Rowland circle grazing incidence spectrometer used as a monochromator for a soft x-ray Manson source in order to calibrate both the source and detectors over the 0.1-1.2 keV spectral range. The originality of the instrument lies on a patented vacuum manipulator which allows the simultaneous boarding of two detectors, one (reference) for measuring the monochromatic radiation and the second to be calibrated. In order to achieve this, the vacuum manipulator is able to interchange, in vacuum, one detector with the other in front of the exit slit of the monochromatizing stage. One purpose of this apparatus was to completely eliminate the intrinsic bremsstrahlung emission of the x-ray diode source and isolate each characteristic line for quantitative detector calibrations. Obtained spectral resolution (Delta lambda/lambda < 10(-2)) and spectral purity (> 98%) fully meet this objective. Initially dimensioned to perform calibration of bulky x-ray cameras unfolded on the Laser MegaJoule Facility, other kinds of detector can be obviously calibrated using this instrument. A brief presentation of the first calibration of an x-ray CCD through its quantum efficiency (QE) measurement is included in this paper as example. Comparison with theoretical model for QE and previous measurements at higher energy are finally presented and discussed.