A randomised comparison of the self-pressurised air-Q (TM) intubating laryngeal airway with the LMA Unique (TM) in children

作者:Jagannathan N*; Sohn L E; Sawardekar A; Shah R; Ryan K; Jagannathan R; Anderson K
来源:Anaesthesia, 2012, 67(9): 973-979.
DOI:10.1111/j.1365-2044.2012.07199.x

摘要

We conducted a randomised trial comparing the self-pressurised air-QTM intubating laryngeal airway (air-Q SP) with the LMA-Unique in 60 children undergoing surgery. Outcomes measured were airway leak pressure, ease and time for insertion, fibreoptic examination, incidence of gastric insufflation and complications. Median (IQR [range]) time to successful device placement was faster with the air-Q SP (12 (1015 [518])) s than with the LMA-Unique (14 (1217 [622]) s; p = 0.05). There were no statistically significant differences between the air-Q SP and LMA-Unique in initial airway leak pressures (16 (1418 [1029]) compared with 18 (1520 [1030]) cmH2O, p = 0.12), an airway leak pressures at 10 min (19 (1622 [1230]) compared with 20 (1622 [1030]) cmH2O, p = 0.81); fibreoptic position, incidence of gastric insufflation, or complications. Both devices provided effective ventilation without the need for airway manipulation. The air-Q SP is an alternative to the LMA-Unique should the clinician prefer a device not requiring cuff monitoring during anaesthesia.