Negative Interference by Rheumatoid Factor of Plasma B-Type Natriuretic Peptide in Chemiluminescent Microparticle Immunoassays

作者:Fan Wen; Xu Lei; Xie Liangcai; Yang Decai; Liu Xuezheng; Zhang Jiajun; Li Yirong*; Yi Cunjian
来源:PLos One, 2014, 9(8): e105304.
DOI:10.1371/journal.pone.0105304

摘要

Background: The chemiluminescent microparticle immunoassay (CMIA) is widely used for the quantitative determination of B-type natriuretic peptide (BNP) in human ethylenediaminetetraacetic acid plasma. Rheumatoid factor (RF) is usually thought to result in a positive interference in immunoassays, but it is not clear whether its presence in plasma can lead to interferences in the CMIA of BNP. Methods: The estimation of BNP recovery was carried out by diluting high-concentration BNP samples with RF-positive or RF-negative plasma at a ratio of 1:9. The diluted samples were then tested using the ARCHITECT i2000 System and ARCHITECT BNP Reagent Kits and the recovery was then calculated. Results: When the RF level ranged from 48 to 1420 IU/mL, the average recovery of BNP was 79.29% and 91.61% in the RF-positive and RF-negative plasma samples, respectively, and was thus significantly lower in the group of RF-positive plasma samples than in the group of RF-negative plasma samples. At a dilution of 1:16, the measured BNP level increased by >36% in six of the seven RF-positive plasma samples. The recovery of BNP increased significantly in the RF-positive plasma samples after pretreatment with IgG-sensitive latex particles. In addition, The BNP recovery was not significantly related to the plasma RF at concentrations ranging from 48 to 2720 IU/mL. Conclusions: Measurement of BNP by CMIA is susceptible to interference from RF leading to predominantly (but not exclusively) lower results. Pretreatment of samples with blocking reagents is advisable prior to the initiation of denying patient's necessary treatment.