Determination of the gamma-secretase inhibitor MK-0752 in human plasma by online extraction and electrospray tandem mass spectrometry (HTLC-ESI-MS/MS)

作者:Bai Feng; Tagen Michael; Colotta Courtney; Miller Laura; Fouladi Maryam; Stewart Clinton F*
来源:Journal of Chromatography B-Analytical Technologies in the Biomedical and Life Sciences, 2010, 878(25): 2348-2352.
DOI:10.1016/j.jchromb.2010.07.019

摘要

A sensitive and rapid HTLC-ESI-MS/MS method with an advanced online sample preparation was developed for determination of the gamma-secretase inhibitor MK-0752 in human plasma using an internal standard. Plasma samples (100 mu l) were diluted and injected directly onto an online extraction column (Cohesive Cyclone MAX 0.5 mm x 50 mm, >30 mu m), the sample matrix was washed out with an aqueous solution, and retained analytes were eluted out and transferred directly to the analytical column (Phenomenex Gemini 3 mu. C18 110A, 50 mm x 2.0 mm at 50 degrees C) for separation using a gradient mobile phase. The eluted analytes were then detected on an API-3000 LC-MS/MS System with ESI and a negative multiple reaction monitoring mode. The monitored ion transitions were m/z 441 -> 175 for MK-0752 and 496 175 for the internal standard. Online extraction recoveries were 81%. The method was validated and was linear in the range of 0.05-50 mu g/mL.. Within-day and between-day precisions were <8.6%, and accuracies were 0.7 and 7.1%. This method was applied to the measurement of plasma MK-0752 levels in a Phase I study of pediatric patients with recurrent or refractory brain tumors.

  • 出版日期2010-9-1