A High-Frequency Linear Ultrasonic Array Utilizing an Interdigitally Bonded 2-2 Piezo-Composite

作者:Cannata Jonathan M*; Williams Jay A; Zhang Lequan; Hu Chang Hong; Shung K Kirk
来源:IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2011, 58(10): 2202-2212.
DOI:10.1109/TUFFC.2011.2070

摘要

This paper describes the development of a high-frequency 256-element linear ultrasonic array utilizing an interdigitally bonded (IB) piezo-composite. Several IB composites were fabricated with different commercial and experimental piezoelectric ceramics and evaluated to determine a suitable formulation for use in high-frequency linear arrays. It was found that the fabricated fine-scale 2-2 IB composites outperformed 1-3 IB composites with identical pillar-and kerf-widths. This result was not expected and lead to the conclusion that dicing damage was likely the cause of the discrepancy. Ultimately, a 2-2 composite fabricated using a fine-grain piezoelectric ceramic was chosen for the array. The composite was manufactured using one IB operation in the azimuth direction to produce approximately 19-mu m-wide pillars separated by 6-mu m-wide kerfs. The array had a 50 mu m (one wavelength in water) azimuth pitch, two matching layers, and 2 mm elevation length focused to 7.3 mm using a polymethylpentene (TPX) lens. The measured pulse-echo center frequency for a representative array element was 28 MHz and -6-dB bandwidth was 61%. The measured single-element transmit -6-dB directivity was estimated to be 50 degrees. The measured insertion loss was 19 dB after compensating for the effects of attenuation and diffraction in the water bath. A fine-wire phantom was used to assess the lateral and axial resolution of the array when paired with a prototype system utilizing a 64-channel analog beamformer. The -6-dB lateral and axial resolutions were estimated to be 125 and 68 mu m, respectively. An anechoic cyst phantom was also imaged to determine the minimum detectable spherical inclusion, and thus the 3-D resolution of the array and beamformer. The minimum anechoic cyst detected was approximately 300 mu m in diameter.

  • 出版日期2011-10