摘要

Purpose It is generally accepted that using a video laryngoscope is associated with an improved visualization of the glottis. However, correctly placing the endotracheal tube might be challenging. Channeled video laryngoscopic blades have an endotracheal tube already pre-loaded, allowing to advance the tube once the glottis is visualized. We hypothesized that use of a channel blade with pre-loaded endotracheal tube results in a faster intubation, compared to a curved Macintosh blade video laryngoscope. Methods After ethical approval and informed consent, patients were randomized to receive endotracheal Intubation with either the King Vision (R) video laryngoscope with curved blade (control) or channeled blade (channeled). Success rate, evaluation of the glottis view (percentage of glottic opening (POGO), Cormack&Lehane (C&L)) and intubating time were evaluated. Results Over a two-month period, a total of 46 patients (control n = 23; channeled n = 23) were examined. The first attempt success rates were comparable between groups (control 100% (23/23) vs. channeled 96% (22/23); p = 0.31). Overall intubation time was significantly shorter with control (median 40 sec; IQR [24-58]), compared to channeled (59 sec [40-74]; p = 0.03). There were no differences in glottis visualization between groups. Conclusion Compared with the King Vision channeled blade, time for tracheal intubation was shorter with the control group using a non-channeled blade. First attempt success and visualization of the glottis were comparable. These data do not support the hypothesis that a channeled blade is superior to a curved video laryngoscopic blade without tube guidance.

  • 出版日期2017-8-31