Microfluidic Amperometric Sensor for Analysis of Nitric Oxide in Whole Blood

作者:Hunter Rebecca A; Privett Benjamin J; Henley W Hampton; Breed Elise R; Liang Zhe; Mittal Rohit; Yoseph Benyam P; McDunn Jonathan E; Burd Eileen M; Coopersmith Craig M; Ramsey J Michael; Schoenfisch Mark H*
来源:Analytical Chemistry, 2013, 85(12): 6066-6072.
DOI:10.1021/ac400932s

摘要

Standard photolithographic techniques and a nitric oxide (NO) selective xerogel polymer were utilized to fabricate an amperometric NO microfluidic sensor with low background noise and the ability to analyze NO levels in small sample volumes (similar to 250 mu L). The sensor exhibited excellent analytical performance in phosphate buffered saline, including a NO sensitivity of 1.4 pA nM(-1), a limit of detection (LOD) of 840 pM, and selectivity over nitrite, ascorbic acid, acetaminophen, uric acid, hydrogen sulfide, ammonium, ammonia, and both protonated and deprotonated peroxynitrite (selectivity coefficients of -5.3, -4.2, -4.0, -5.0, -6.0, -5.8, -3.8, -1.5, and -4.0, respectively). To demonstrate the utility of the microfluidic NO sensor for biomedical analysis, the device was used to monitor changes in blood NO levels during the onset of sepsis in a murine pneumonia model.

  • 出版日期2013-6-18