摘要

Laser scanning cytometry (LSC) is emerging as a clinical tool. In one application a "Clatch" slide, named after the inventor, is used in conjunction with LSC for cell surface marker immunophenotyping of patient samples. The slide requires time consuming and laborious pipetting steps, making a test tedious and prone to handling errors. The Clatch slide also uses a significant number of cells, limiting the number of analyses on paucicellular samples. This paper presents an automated microfluidic system consisting of a control circuit, a microfluidic system, and an aluminum frame, capable of performing immunophenotyping procedures. This prototype system reduces 36 pipetting steps to 1, reduces the amount of cell sample from 180 mu L to 56 mu L, and shortens the time used by technicians.

  • 出版日期2011-4