Inconsistencies in the ITS-90 interpolations above 0.01 A degrees C have been investigated. Thirty-five standard platinum resistance thermometers are used to determine the magnitude of inconsistencies in this range with the exception of the overlap region of the water-zinc and water-aluminum sub-ranges. The results show inconsistencies that are less 4.1 mK. Simpler representations of ten of the fifteen inconsistency functions above 0.01 A degrees C are presented by means of factoring and extremum analysis.
- 出版日期2011-1
- 单位中国计量科学研究院; 河北省气象科学研究所