Axial or Helical? Considerations for wide collimation CT scanners capable of volumetric imaging in both modes

作者:Lambert Jack W*; Phillips Elizabeth D; Villanueva Meyer Javier E; Nardo Lorenzo; Facchetti Luca; Gould Robert G
来源:Medical Physics, 2017, 44(11): 5718-5725.
DOI:10.1002/mp.12525

摘要

PurposeTo determine whether axial or helical mode is more appropriate for a 16cm collimation CT scanner capable of step-and-shoot volumetric axial coverage, in terms of radiation dose, image quality, and scan duration. MethodsAll scans were performed with a Revolution CT (GE Healthcare) operating at 120kV and 100mAs. Using calibrated optically stimulated luminescence detectors, radiation dose along the axial scan profile was evaluated at the isocenter, including the overlap region between two axial sections. This overlap region measures 3 cm in the z-axis at the isocenter and is required to obtain sufficient projection data from the relatively large cone-beam angles. Using an image quality phantom (Gammex Model 464), spatial resolution, CT number uniformity, image noise, and low contrast detectability (LCD) were evaluated under five different conditions: in the middle of a helical acquisition, in the middle of a 16cm axial section, at both ends of an axial section and in the overlap region between two axial sections. Scan durations and dose length products (DLP) were recorded for prescribed scan lengths of 2.5-100cm. ResultsThe overlap region between two axial sections received a dose 83% higher than the single-exposure region at the isocenter. Within a single axial section, the dose at the anode end was 37% less than at the cathode end due to the anode heel effect. Image noise ranged from a low of 13HU for the cathode end of an axial section up to 14.7HU for the anode end (P<0.001). The LCD was at lower at the anode end of the axial section compared to both the cathode end (P<0.05) and the overlap location (P<0.02). The spatial resolution and CT number uniformity were consistent among all conditions. Scan durations were shorter (0.28s) for the axial mode compared to the helical mode at scan lengths 16cm, and longer at scan lengths 16cm where more than one table position was required, up to a difference of 13.9s for a the 100cm scan length (3.8s for helical compared to 17.6s for axial). DLPs were consistent between scan modes; slightly lower in axial mode at shorter scan lengths due to helical overranging, and slightly higher in axial mode at longer scan lengths due to the axial overlap regions. ConclusionsTo ensure the most consistent radiation dose and image quality along the scan length, we recommend helical mode for scans longer than the 16cm coverage of a single axial section. For scan lengths 16cm, axial scanning is the most practical option, with a shorter scan duration and higher dose efficiency.

  • 出版日期2017-11