摘要

In this paper, an improvement of the temperature programmed desorption (TPD) technique is introduced, which facilitates fully automated sampling of TPD spectra with excellent reproducibility, especially useful for extremely low coverages. By averaging many sampled TPD spectra, the range of the TPD technique can be extended toward lower coverages, as well as the quality of the spectra can be improved. This allows for easy extraction of information about the adsorbate-surface bond. A state of the art TPD apparatus with a two chamber setup and a high quality quadrupole mass spectrometer was extended by automated components. These are an automated gas dosing system, ensuring precise dosing of gas, combined with a motor driven sample manipulation unit and a liquid nitrogen cryostat with automatic refilling. In addition all components were controlled by a computer. A large number of TPD cycles could be sampled without the need of interaction of an operator. Here, it is shown for up to more than 400 TPD cycles. This opens a wide range of new interesting applications for the TPD technique, especially in the limit of zero coverage. Here, basic experiments on well known adsorbate systems are shown to view the ability and limit of this approach.

  • 出版日期2010-3
  • 单位AREVA NP