A wireless power transmission system for implantable devices in freely moving rodents

作者:Eom Kyungsik; Jeong Joonsoo; Lee Tae Hyung; Kim Jinhyung; Kim Junghoon; Lee Sung Eun; Kim Sung June*
来源:Medical & Biological Engineering & Computing, 2014, 52(8): 639-651.
DOI:10.1007/s11517-014-1169-3

摘要

Reliable wireless power delivery for implantable devices in animals is highly desired for safe and effective experimental use. Batteries require frequent replacement; wired connections are inconvenient and unsafe, and short-distance inductive coupling requires the attachment of an exterior transmitter to the animal's body. In this article, we propose a solution by which animals with implantable devices can move freely without attachments. Power is transmitted using coils attached to the animal's cage and is received by a receiver coil implanted in the animal. For a three-dimensionally uniform delivery of power, we designed a columnar dual-transmitter coil configuration. A resonator-based inductive link was adopted for efficient long-range power delivery, and we used a novel biocompatible liquid crystal polymer substrate as the implantable receiver device. Using this wireless power delivery system, we obtain an average power transfer efficiency of 15.2 % (minimum efficiency of 10 % and a standard deviation of 2.6) within a cage of 15 x 20 x 15 cm(3).

  • 出版日期2014-8