Usefulness of direct-conversion flat-panel detector system as a quality assurance tool for high-dose-rate Ir-192 source

作者:Miyahara Yoshinori*; Kitagaki Hajime; Nishimura Tomonori; Itakura Kanae; Takahashi Shinobu; Yokokawa Masaki; Uchida Nobue; Inomata Taisuke
来源:Journal of Applied Clinical Medical Physics, 2015, 16(1): 121-129.

摘要

The routine quality assurance (QA) procedure for a high-dose-rate (HDR) Ir-192 radioactive source is an important task to provide appropriate brachytherapy. Traditionally, it has been difficult to obtain good quality images using the Ir-192 source due to irradiation from the high-energy gamma rays. However, a direct-conversion flat-panel detector (d-FPD) has made it possible to confirm the localization and configuration of the Ir-192 source. The purpose of the present study was to evaluate positional and temporal accuracy of the Ir-192 source using a d-FPD system, and the usefulness of d-FPD as a QA tool. As a weekly verification of source positional accuracy test, we obtained Ir-192 core imaging by single-shot radiography for three different positions (1300/1400/1500 mm) of a check ruler. To acquire images for measurement of the Ir-192 source movement distance with varying interval steps (2.5/5.0/10.0 mm) and temporal accuracy, we used the high-speed image acquisition technique and digital subtraction. For accuracy of the Ir-192 source dwell time, sequential images were obtained using various dwell times ranging from 0.5 to 30.0 sec, and the acquired number of image frames was assessed. Analysis of the data was performed using the measurement analysis function of the d-FPD system. Although there were slight weekly variations in source positional accuracy, the measured positional errors were less than 1.0 mm. For source temporal accuracy, the temporal errors were less than 1.0%, and the correlation between acquired frames and programmed time showed excellent linearity (R-2 = 1). All Ir-192 core images were acquired clearly without image halation, and the data were obtained quantitatively. All data were successfully stored in the picture archiving and communication system (PACS) for time-series analysis. The d-FPD is considered useful as the QA tool for the Ir-192 source.

  • 出版日期2015