摘要

PADC detectors are widely used as air radon concentration measurement devices and the typical procedure that a Radon Service Laboratory uses to manage the detectors provides a calibration phase followed by the field measurement. The calibration is performed in a reference radon concentration atmosphere, using high radon concentration values in order to achieve typical exposure values of few MBq h m(-3) with an exposure time of few days. On the other hand the field measurement is characterized by long term exposures lasting up to six months and by radon concentrations that are quite lower than the ones used in the calibration.
The aim of this study is to check whether the calibration procedure is actually representative of a field measurement, or, in other words, whether and how much ageing or fading can affect the calibration factor.
We found that the ageing and fading effect can produce a decrease in the detector sensitivity leading to an underestimation of the radon concentration up to 40% for exposure lasting few months. An important issue is that both ageing and fading can be ascribed to a decrease in the track etching velocity V(t).
In the paper we also provide an algorithm to compensate for the sensitivity variation due to fading/ageing effect.