A New N-Heterocyclic Carbene Ionophore in Plasticizer-free Polypyrrole Membrane for Determining Ag+ in Tap Water

作者:Said Nur Rahimah; Rezayi Majid; Narimani Leila; Al Mohammed Nassir N; Manan Ninie Suhana Abdul; Alias Yatimah
来源:Electrochimica Acta, 2016, 197: 10-22.
DOI:10.1016/j.electacta.2016.02.173

摘要

A bis[1-benzyl-benzimidazoliumethyl]-4-methylbenzenesufonamide bromide ligand (NHCL) as a new type of ionophore was successfully synthesized for the first time and used for the fabrication of Ag+ ion selective electrode as a new type of ionophore, based on the self-plasticizing method. The ionophore was characterized using CHN elemental analysis, FT-IR, H-1 NMR, C-13 NMR, UV-vis spectroscopic techniques and X-ray single crystal diffraction. The interection between NHCL and Ag+ cation was studied spectrophotometerically and the stoichiometry of the complex (1: 2) NHCL: Ag+ in DMSO solvent was determined. The potentiometric response of the electrode toward Ag+ cation was linear from 2.5 x 10(-6) to 1.0 x 10(-1) M with detection limit of 2.00 x 10(-6) M and a Nernstian slope of 58.48 +/- 0.75 mV decade(-1). The sensor exhibited a response time of about 20 seconds at room temperature and working pH range between 3.5-9.0. The electrode showed a good selectivity towards Ag+ cations in comparison with other soft and hard metals, while most of these metal ions did not show significant interference (K-Ag+,M(pot)). The proposed electrode also applied in the direct determination of Ag+ cations in tap water and standard samples with sensible accuracy and precision.

  • 出版日期2016-4-10