Anchor placement and subsequent movement in a mesh kit with self-fixating tips: 6-month follow-up of a prospective cohort

作者:Brennand E A*; Bhayana D; Tang S; Birch C; Murphy M; Cenaiko D; Ross S; Robert M
来源:BJOG: An International Journal of Obstetrics and Gynaecology , 2014, 121(5): 634-640.
DOI:10.1111/1471-0528.12536

摘要

ObjectiveTo describe the initial placement of Elevate single-incision mesh kit device tips relative to the sacrospinous ligament, and to measure tip movement over a 6-month period from initial placement. DesignProspective cohort. SettingTertiary care urogynaecology centre in Calgary, Canada. PopulationWomen electing for surgical management of anterior vaginal wall prolapse. MethodsTen women underwent anterior prolapse repair using the Elevate single-incision mesh kit with a metallic fiducial marker attached to the tips of the surgical device. Women were imaged by magnetic resonance imaging (MRI) within 48hours of surgery, and again 6months later to investigate the position of the device tips and change 4mm over the 6-month postoperative period. Main outcome measure(s)Position of self-fixating tips within 48 hours of surgery, and at six months post-operative. ResultsAnchor insertion was directly into the sacrospinous ligament in 10 of 20 insertion points (50%, 95% CI 27-73%). Movement was most often noted in the cranial-caudal direction: a change in location of 4mm was observed for 8/20 anchors (40%, 95% CI 19-64%). Cranial-caudal movement was observed less frequently among sacrospinous anchors than among anchors inserted into other pelvic structures (1/10 versus 7/10, P=0.020, difference in proportion -60%, 95% CI -94 to -26%). PFDI-20 scores improved statistically significantly by 6months (P=0.008, mean change -62.9%, 95% CI -105.1 to -20.7%), but PFIQ-7 scores did not change statistically significantly over the same time period (P=0.523, mean change -12.4%, 95% CI -54.5 to 29.8%). ConclusionsThe novel self-fixating anchoring tips of this single-incision mesh kit do not reliably anchor into the sacrospinous ligament. The tips have been shown to move with time, although not all cases of anchor movement were associated with recurrent prolapse.

  • 出版日期2014-4