Does Rebound Tonometry Probe Misalignment Modify Intraocular Pressure Measurements in Human Eyes?

作者:Beasley Ian G; Laughton Deborah S; Coldrick Benjamin J; Drew Thomas E; Sallah Marium; Davies Leon N*
来源:Journal of Ophthalmology, 2013, 2013: 791084.
DOI:10.1155/2013/791084

摘要

Purpose. To examine the influence of positional misalignments on intraocular pressure (IOP) measurement with a rebound tonometer. Methods. Using the iCare rebound tonometer, IOP readings were taken from the right eye of 36 healthy subjects at the central corneal apex (CC) and compared to IOPmeasures using the Goldmann applanation tonometer (GAT). Using a bespoke rig, iCare IOP readings were also taken 2 mm laterally from CC, both nasally and temporally, along with angular deviations of 5 and 10 degrees, both nasally and temporally to the visual axis. Results. Mean IOP +/- SD, asmeasured byGAT, was 14.7 +/- 2.5 mmHg versus iCare tonometer readings of 17.4 +/- 3.6mmHg at CC, representing an iCare IOP overestimation of 2.7 +/- 2.8mm Hg (P %26lt; 0.001), which increased at higher average IOPs. IOP at CC using the iCare tonometer was not significantly different to values at lateral displacements. IOP wasmarginally underestimated with angular deviation of the probe but only reaching significance at 10 degrees nasally. Conclusions. As shown previously, the iCare tonometer overestimates IOP compared to GAT. However, IOP measurement in normal, healthy subjects using the iCare rebound tonometer appears insensitive to misalignments. An IOP underestimation of %26lt; 1mmHg with the probe deviated 10 degrees nasally reached statistical but not clinical significance levels.

  • 出版日期2013