Analytical performance of glucometers in a tertiary care hospital

作者:Higioka Angela Somavilla; Messores Rudolf Oliveira Renata Cristina; Martinello Flavia; Bazzo Maria Luiza*
来源:Accreditation and Quality Assurance, 2016, 21(2): 143-149.
DOI:10.1007/s00769-015-1188-2

摘要

Point-of-care (POC) testing of glucose (glucometers) represents a convenient alternative to monitor glycemia since the measurement procedure is performed without delay after sampling of the capillary blood, thereby avoiding the metabolism by the blood cells of glucose present in plasma. Likely because of sample instability, there is no proficiency test provider in Brazil for this type of POC sample. In this context, this study aimed to evaluate the analytical performance of glucometers used in a tertiary care hospital. The glucometers used were the Accu-Chek Performa(A (R)) model from Roche Diagnostics, which use the principle of amperometry. The reference method was the reaction with modified hexokinase/glucose-6-phosphate in a Dimension(A (R)) device. The stability evaluation of the control samples showed that it can be performed up to 90 min after the collection of whole blood samples. In the two rounds performed, only one result of the 17 glucometers evaluated was out of the threshold of two standard deviation. Thus, this method for control of glucometers met the expectations and enabled comparing the glucometers in a hospital. Given the current quality guidelines, daily internal quality control of glucometers is recommended, besides at least two annual comparisons between the results of the glucometers and the reference method and one EQA every 3 months.

  • 出版日期2016-4