A STUDY IN EUROPE OF PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY: PROTOCOL DEVELOPMENT AND FINDINGS

作者:Gfirtner H; Kaplanis P A; Moores B M*; Schneider P; Vassileva J
来源:Radiation Protection Dosimetry, 2010, 139(1-3): 380-387.
DOI:10.1093/rpd/ncq025

摘要

In a multi-centre study more than 600 patient measurements of patients receiving thorax radiography were performed in five European cities. Participating centres were Fulda, Liverpool, Nicosia, Passau and Sofia. The dose quantities measured were the air kerma-area product (AKAP), incident air kerma and entrance surface air kerma (ESAK). In addition, for each patient sex, age, weight, height, focus-to-film distance, focus-to-skin distance, patient thickness, kV, mAs, field size and the nominal film speed were registered. Different X-ray machines were used in the participating centres-Thoravision, flat panel detector and film screen combinations in two centres (analogue system). The tube voltages employed varied between 60 and 150 kV and the nominal film speed between 200 and 400/800. All mean dose values (ESAK and AKAP) for the different centres showed a different value for female and male populations. The differences were up to 100 % and always higher for the mean AKAP. For a thorax posteroanterior examination, the mean ESAK varied between 0.06 and 0.46 mGy and the mean AKAP varied between 60 and 690 mGy cm(2). The differences in the results obtained as well as the methodologies for multinational, multi-centre studies will be discussed. Future perspectives for this type of study within the framework of radiation protection and quality assurance in Europe will also be discussed as well as the role and function of multi-national radiological data sets including patient dose values.