A High-Precision Passive Air Sampler for Gaseous Mercury

作者:McLagan David S; Mitchell Carl P J; Huang Haiyong; Lei Ying Duan; Cole Amanda S; Steffen Alexandra; Hung Hayley; Wania Frank*
来源:Environmental Science and Technology Letters, 2016, 3(1): 24-29.
DOI:10.1021/acs.estlett.5b00319

摘要

Passive air samplers (PASs) provide an opportunity to improve the spatial range and resolution of gaseous mercury (Hg) measurements. Here, we propose a sampler design that combines a sulfur-impregnated activated carbon sorbent, a Radiello diffusive barrier, and a protective shield for outdoor deployments. The amount of gaseous Hg taken up by the sampler increased linearly with time for both an 11-week indoor (r(2) = 0.990) and 12-month outdoor (r(2) = 0.996) deployment, yielding sampling rates of 0.158 +/- 0.008 m(3) day(-1) indoors and 0.121 +/- 0.005 m(3) day(-1) outdoors. These sampling rates are close to modeled estimates of 0.166 m(3) day(-1) indoors and 0.129 m(3) day(-1) outdoors. Replicate precision is better than for all previous PASs for gaseous Hg, especially during outdoor deployments (2 +/- 1.3%). Such precision is essential for discriminating the relatively small concentration variations occurring at background sites. Deployment times for obtaining reliable time-averaged atmospheric gaseous Hg concentrations range from a week to at least one year.

  • 出版日期2016-1