Multielement analytical procedure coupling INAA, ICP-MS and ICP-AES: Application to the determination of major and trace elements in sediment samples of the Bouregreg river (Morocco)

作者:Bounouira H*; Choukri A; Cherkaoui R; Gaudry A; Delmas R; Mariet C; Hakam O K; Chakiri S
来源:Journal of Radioanalytical and Nuclear Chemistry, 2008, 278(1): 65-79.
DOI:10.1007/s10967-007-6937-9

摘要

Instrumental neutron activation analysis (INAA), inductively coupled plasma-mass spectrometry (ICP-MS) and inductively coupled plasma-atomic emission spectrometry (ICP-AES) were used for the determination of major and trace elements in sediment samples of the Bouregreg river (Morocco). The reliability of the results was checked, by using IAEA Soil-7 certified reference material. Results obtained by the three techniques were compared to control digestions efficiencies. A general good agreement was found between INAA and both ICP-MS and ICP-AES after alkaline fusion (ICPf). The ICP-MS technique used after acid attack (ICPa) was satisfactory for a few elements. A principal component analysis (PCA) has been used for analyzing the variability of concentrations, and defining the most influential sites with respect to the general variation trends. Three groups of elements could be distinguished. For these groups a normalization of concentrations to the central element concentration (that means Mn, Si or Al) is proposed.

  • 出版日期2008-10
  • 单位中国地震局