摘要

A simple, rapid and sensitive spectrofluorimetric method was developed for the determination of di-syston, ethion and phorate in environmental water samples. The procedure is based on the oxidation of these pesticides with cerium (IV) to produce cerium (III), and its fluorescence was monitored at 368 +/- A 3 nm after excitation at 257 +/- A 3 nm. The variables effecting oxidation of each pesticide were studied and optimized. Under the experimental conditions used, the calibration graphs were linear over the range 0.2-15, 0.1-13, 0.1-13 ng mL(-1) for di-syston, ethion and phorate, respectively. The limit of detection and quantification were in the range 0.034-0.096 and 0.112-0.316 ng mL(-1), respectively. Intra- and inter-day assay precisions, expressed as the relative standard deviation (RSD), were lower than 5.2 % and 6.7 %, respectively. Good recoveries in the range 86 %-108 % were obtained for spiked water samples. The proposed method was applied to the determination of studied pesticides in environmental water samples.