摘要

N-Phenylanthranilic acid was used as a complexing agent for determination of uranium(VI) by adsorptive cathodic stripping voltammetry. Under the optimal experimental conditions of the experimental parameters, the peak current was proportional to the concentration of U(VI) in the range 0.75-30 ng mL(-1) and the detection limit was 0.036 ng mL(-1). The influence of possible interferences was investigated. The method was applied for determination of uranium in waste water from uranium conversion facility and natural water samples. Application of the method for simultaneous determination of U(VI) and Cu(II) showed that these ions could be simultaneously determined in a single scan at relatively wide concentration range.

  • 出版日期2014-7