摘要

Aims: To investigate the diagnostic quality of different quality, individually calibrated ink-jet printers for the very challenging dental radiographic task of approximal carious lesion detection.
Materials and methods: A test-pattern evaluating resolution, contrast and homogeneity of the ink-jet prints was developed. 50 standardized dental radiographs each showing two neighbouring teeth in natural contact were printed on glossy paper with calibrated, randomly selected ink-jet printers (Canon S520 and iP4500, Epson Stylus Photo R2400). Printing size equalled the viewing size on a 17 '' cathode-ray-tube monitor daily quality-tested according to German regulations. The true caries status was determined from serial sectioning and microscopic evaluation. 16 experienced observers evaluated the radiographs on a five-point confidence scale on all prints plus the viewing monitor with respect to the visibility of a carious lesion. A non-parametric Receiver-Operating Characteristics (ROC-) analysis was performed explicitly designed for the evaluation of readings stemming from identical samples but different modality. Significant differences are expressed by a critical ratio z exceeding +/- 2. Diagnostic accuracy was determined by the area (Az) underneath the ROC- curves.
Results: Average Az-values ranged between 0.62 (S520 and R2400) and 0.64 (monitor, iP4500), with no significant difference between modalities (P = 0.172). Neither significant (range mean z:-0.40 (S520) and -0.11 (iP4500)) nor clinically relevant differences were found between printers and viewing monitor.
Conclusions: Our results for a challenging task in dental radiography indicate that calibrated, off-the-shelf ink-jet printers are able to reproduce (dental) radiographs at quality levels sufficient for radiographic diagnosis in a typical dental working environment.

  • 出版日期2011-8