Accuracy of portable devices in measuring peak cough flow

作者:Kulnik Stefan Tino*; MacBean Victoria; Birring Surinder Singh; Moxham John; Rafferty Gerrard Francis; Kalra Lalit
来源:Physiological Measurement, 2015, 36(2): 243-257.
DOI:10.1088/0967-3334/36/2/243

摘要

Peak cough flow (PCF) measurements can be used as indicators of cough effectiveness. Portable peak flow meters and spirometers have been used to measure PCF, but little is known about their accuracy compared to pneumotachograph systems. The aim of this study was to compare the accuracy of four portable devices (Mini-Wright and Assess peak flow meters, SpiroUSB and Microlab spirometers) in measuring PCF with a calibrated laboratory based pneumotachograph system. Twenty healthy volunteers (mean (SD) age 45 (16) years) coughed through a pneumotachograph connected in series with each portable device in turn, and the differences in PCF readings were analysed. In addition, mechanically generated flow waves of constant peak flow were delivered through each device both independently and when connected in series with the pneumotachograph. Agreement between PCF readings obtained with the pneumotachograph and the portable devices was poor. Peak flow readings were on average lower by approximately 50 L min(-1) when measured using the portable devices; 95% limits of agreement spanned approximately 150 L min(-1). The findings highlight the potential for inaccuracy when using portable devices for the measurement of PCF. Depending on the measurement instrument used, absolute values of PCF reported in the literature may not be directly comparable.

  • 出版日期2015-2